Input
BCD display
Measurement mode
Parallel BCD - 24 bit
Binary - 20 bit
Serial BCD - 4 data/6 Strobe
Serial BCD - 8 data/3 Strobe
Serial BCD - 13 data/2 Strobe
Serial BCD - 4 data/1 Strobe/3 Segments
Address
up to 8 projections
BCD - transformer tapping display
Range
5…24 VDC
10…60 VDC
90…130 VDC
190…250 VDC
Number of tapping leads
24 + 1 signalization
(on request 27)
Input resistence
5,5 kOhm/V
Output
relay BIN/BCD
5 relays (250 VAC/50 VDC, 3 A)
Relay mode
Projection
Display range
-99999…999999, single-coloured 14-segment LED
Display colour
red, green
Signalizing LED
1, 2, 3, 4
|
Limits L1, L2, L3, L4 (red)
|
M
|
Min./Max. value (green)
|
Description
the last two digits of the display can be used to display measured variables (only 14 mm display)
Displayed values
Kan A
|
Value from channel A
|
Fil. A
|
Value from channel A after digital filter adjustment
|
Mat. Fn.
|
Value of mathematical function
|
Max.
|
Maximum value
|
Min.
|
Minimum value
|
Settings
Control
The device is controlled and set by five buttons on the front panel
Menu
PROFI
|
Full device settings with password access
|
Instrument accuracy and functionality
Temperature coefficient
50 ppm/°C
Digital filters
floating mean
|
from 2…30 measurements
|
exponential mean
|
from 2…100 measurements
|
rounding
|
setting projection step for display
|
Functions
Min/max value
|
registration of min./max. value achieved during measurement
|
Math. functions
polynom, 1/x, logarithm, exponential, power, root, sin x
Watch-dog
reset after 1,2 s
Calibration
at 25°C and 40 % r.h.
Comparators
Type
Digital, menu adjustable, contact connect. < 30 ms
Mode
Hysteresis
|
switch. point, hysteresis band "Mez ± 1/2 Hys." and time (0 ... 99.9 s) determ. switching delay
|
Dose
|
period, its multiples and time (0 ... 99.9 s), during which the output is active
|
Output
1x relay, switch-over contact (250 VAC/30 VDC, 3 A)
2x relay, switch-over contact (250 VAC/30 VDC, 3 A)
1x relay, switch on contact (250 VAC/30 VDC, 3A)
2x relay, switch on contact (250 VAC/30 VDC, 3 A)
3x relay, switch on contact (250 VAC/30 VDC, 3A)
Data outputs
Data format
8 bits + no parity + 1 stop bit (ASCII)
7 bits + even parity + 1 stop bit (Messbus)
RS 485
isolated, addressing max. 31 instruments
Analogue outputs
Type
Isolated, programmable with resolution of max.10,000 steps, type and range are menu selectable
Temperature coefficient
50 ppm/°C
Nonlinearity
±0,2 % of range
Rate
response to value change < 40 ms
Ranges
0...2 V, 0...5 V, 0...10 V, 0...5 mA, 0...20 mA, 4...20 mA
Compensation
600 Ohm/12 V
Sensor excitation
Adjustable
5…24 VDC/max. 1,2 W
Power supply
Power supply
9…50 V AC/DC, ±10 %, PF > 0,4, I < 40 A/1 ms, isolated
80…250 V AC/DC, ±10 %, PF > 0,4, I < 40 A/1 ms, isolated
Current draw
< 6,5 W/6,0 VA
Power supply protection
Power supply is protected by a fuse inside the instrument
Operating conditions
Connection
connector terminal board, cable section < 2,5 mm2
Stabilization period
within 15 minutes after switch on
Operating temperature
-20°…60°C
Storage temperature
-20°…80°C
Protection class
IP64 (front panel only)
El. safety
EN 61010-1, A2
Dielectric strength
4 kVAC within 1 min. between pow. supply and input
4 kVAC within 1 min. between pow. supply and data/anal. output
4 kVAC within 1 min. between input and relay output
2,5 kVAC within 1 min.between input and data/anal. output
Insulation resistance
for polution degree II, measuring cat. III.
power supply
|
670 V (ZI), 300 V (Di)*
|
input, output, PN
|
300 V (ZI), 150 V (Di)*
|
Mechanical properties
Material
Noryl GFN2 SE1, non-inflammable UL 94 V-I
Dimensions
96 x 48 x 154 mm (w x h x d)
Panel cutout
90,5 x 45 mm (w x h)
Images
Dimensions and installation
{"ds01-6949-2023.1-en":"10167","ds01-6949-2023.1-cs":"10168","ds01-6949-2020.1-en":"11081","ds01-6949-2020.1-fr":"11082","ds01-6949-2020.1-ru":"11083","ds01-6949-2020.1-cs":"11098","ds01-6949-2019.1-en":"11078","ds01-6949-2019.1-fr":"11079","ds01-6949-2019.1-ru":"11080","ds01-6949-2019.1-cs":"11097","ds01-6949-2017.1-en":"11076","ds01-6949-2017.1-ru":"11077","ds01-6949-2017.1-cs":"11096","ds01-6949-2015.1-en":"11073","ds01-6949-2015.1-fr":"11074","ds01-6949-2015.1-ru":"11075","ds01-6949-2014.2-en":"11070","ds01-6949-2014.2-fr":"11071","ds01-6949-2014.2-ru":"11072","ds01-6949-2014.2-cs":"11094","ds01-6949-2013.2-en":"11067","ds01-6949-2013.2-fr":"11068","ds01-6949-2013.2-ru":"11069","ds01-6949-2013.2-cs":"11093","ds01-6949-2012.2-en":"11064","ds01-6949-2012.2-fr":"11065","ds01-6949-2012.2-ru":"11066","ds01-6949-2012.2-cs":"11092","ds01-6949-2010.1-en":"11061","ds01-6949-2010.1-fr":"11062","ds01-6949-2010.1-ru":"11063","ds01-6949-2010.1-cs":"11091","ds01-6949-2008.1-en":"11059","ds01-6949-2008.1-ru":"11060","ds01-6949-2008.1-cs":"11089","ds01-6949-2006.2-en":"11058","ds01-6949-2006.2-cs":"11087","ds01-6949-2005.2-ru":"11057","ds01-6949-2003.3-en":"11055","ds01-6949-2003.3-ru":"11056","ds01-6949-2003.3-cs":"11085","ds01-6949-2002.1-en":"11053","ds01-6949-2002.1-ru":"11054","ds01-6949-2002.1-cs":"11084","ds01-6949-2016.1-cs":"11095","ds01-6949-2009.1-cs":"11090","ds01-6949-2007.1-cs":"11088","ds01-6949-2005.1-cs":"11086","man08-6950-2017.5-en":"3671","man09-6950-2016.4.1-cs":"3676","man010-6950-2003.4-cs":"3678","crt02-6952-2022-en":"13197","crt02-6952-2022-cs":"13198","crt03-6952-2022-en":"12969","crt03-6952-2022-cs":"12970","sw-6954-OM621BCD 2-cs":"3693"}
ISOLATED TRANSDUCER • USB < > OM INSTRUMENTS
Small compact transducer OM LINK USB with cable ensures safe and comfortable galvanically isolated connection between your computer and ORBIT MERRET instruments.