Input
AC
Input frequency
0…400 Hz
for amplitude from 8 V
Measured variables
Voltage (Vrms)
Current (Irms)
External inputs
Number
1 input, on contact
Functions
OFF
|
No function
|
HOLD
|
Stopping measurement (option: Display - Analog output - Limits)
|
TARA
|
Tare activation
|
Projection
Display range
0…1999, single-coloured 7-segment LED
Display colour
red, green
Signalizing LED
1
|
Limit L1 (red)
|
T
|
Tare (green)
|
Decimal point
menu adjustable - fixed or floating
Display brightness
menu adjustable, automatically
Settings
Control
The device is controlled and set by five buttons located behind the the rear lid
Selected functions can also be controlled via external inputs
Menu
LIGHT
|
contains only items necessary for basic device settings with password access
|
PROFI
|
Full device settings with password access
|
USER
|
User menu of any item, selected from LIGHT/PROFI. Access without password.
|
OM Link
Company communication interface for operation, setting and update of device firmware
OM Link program is also designed for visualization and filing of meas. values from more instruments
Instrument accuracy and functionality
Temperature coefficient
50 ppm/°C
Accuracy
0,3 % of range + 1 digit
(for projection 0…1999 and 5 measur./s)
Rate
0,5/1,2/2,5/5 measur./s
Overload capacity
2x; 10x (t < 30 ms) - not for > 250 V and 5 A
Linearization
linear interpolation in 25 points (only via OM Link)
Digital filters
exponential mean
|
from 2…100 measurements
|
rounding
|
setting projection step for display
|
Functions
Tare
|
designed to reset display on non-zero input signal
|
Watch-dog
reset after 500 ms
Calibration
at 25°C and 40 % r.h.
Comparators
Type
Digital, menu adjustable, contact connect. < 50 ms
Mode
Hysteresis
|
switch. point, hysteresis band "Mez ± 1/2 Hys." and time (0 ... 99.9 s) determ. switching delay
|
Output
1x relay, switch on contact (250 VAC/30 VDC, 3A)
1x open collector (30 VDC/100 mA)
Power supply
Power supply
10…30 VDC/24 VAC, ±10 %, PF > 0,4, I < 45 A/1 ms, isolated
Current draw
< 1,8 W/1,9 VA
Operating conditions
Connection
terminal board, cable section < 1,5 mm2
Stabilization period
within 15 minutes after switch on
Operating temperature
-20°…60°C
Storage temperature
-20°…80°C
Protection class
IP 65 (front panel only)
El. safety
EN 61010-1, A2
Dielectric strength
2,5 kVAC within 1 min. between pow. supply and input
4 kVAC within 1 min. between input and relay output
Insulation resistance
for polution degree II, measuring cat. III.
power supply
|
300 V (ZI)*
|
input, output
|
300 V (ZI)*
|
Mechanical properties
Material
Polycarbonate, non-inflammable UL 94 V-0
Dimensions
96 x 48 x 30 mm (w x h x d)
Panel cutout
92 x 44 mm (w x h)
Images
Dimensions and installation
{"ds01-6949-2015.1-en":"11691","ds01-6949-2015.1-fr":"11692","ds01-6949-2015.1-ru":"11693","ds01-6949-2015.1-cs":"11704","ds01-6949-2014.2-en":"11688","ds01-6949-2014.2-fr":"11689","ds01-6949-2014.2-ru":"11690","ds01-6949-2014.2-cs":"11703","ds01-6949-2014.1-en":"11685","ds01-6949-2014.1-fr":"11686","ds01-6949-2014.1-ru":"11687","ds01-6949-2014.1-cs":"11702","ds01-6949-2013.2-en":"11682","ds01-6949-2013.2-fr":"11683","ds01-6949-2013.2-ru":"11684","ds01-6949-2013.2-cs":"11701","ds01-6949-2013.1-ru":"11681","ds01-6949-2012.2-en":"11678","ds01-6949-2012.2-fr":"11679","ds01-6949-2012.2-ru":"11680","ds01-6949-2012.2-cs":"11700","ds01-6949-2010.1-en":"11676","ds01-6949-2010.1-fr":"11677","ds01-6949-2010.1-cs":"11698","ds01-6949-2023.1-en":"10201","ds01-6949-2023.1-cs":"10202","ds01-6949-2020.1-en":"11673","ds01-6949-2020.1-fr":"11674","ds01-6949-2020.1-ru":"11675","ds01-6949-2020.1-cs":"11697","ds01-6949-2019.1-en":"11670","ds01-6949-2019.1-fr":"11671","ds01-6949-2019.1-ru":"11672","ds01-6949-2019.1-cs":"11696","ds01-6949-2017.1-en":"11668","ds01-6949-2017.1-ru":"11669","ds01-6949-2017.1-cs":"11695","ds01-6949-2012.1-cs":"11699","ds01-6949-2016.1-cs":"11694","man01-6950-2012.1.0-en":"5563","man01-6950-2012.1.0-cs":"5561","crt02-6952-2022-en":"13273","crt02-6952-2022-cs":"13274","crt03-6952-2022-en":"13079","crt03-6952-2022-cs":"13080","fw01-6953-2012-xx":"5562"}
ISOLATED TRANSDUCER • USB < > OM INSTRUMENTS
Small compact transducer OM LINK USB with cable ensures safe and comfortable galvanically isolated connection between your computer and ORBIT MERRET instruments.