Input
DC
Range
menu selectable
±60 mV
> 100 MOhm
Input U
±150 mV
> 100 MOhm
Input U
±300 mV
> 100 MOhm
Input U
±1200 mV
> 100 MOhm
Input U
OHM
Range
selectable in the menu with automatic change of the range
0…100 Ohm
0…1 kOhm
0…10 kOhm
0…100 kOhm
Connection
2, 3 or 4-wire
RTD
Type
menu selectable
EU > 100/500/1 000 Ohm s 3 850 ppm/°C
-50°…450°C
RU > 50 Ohm s 3 910 ppm/°C
-200°…1 100°C
RU > 100 Ohm s 3 910 ppm/°C
-200°…450°C
US > 100 Ohm s 3 920 ppm/°C
-50°…450°C
Connection
2, 3 or 4-wire
Ni
Type
menu selectable
Ni 1 000/10 000 s 5 000 ppm/°C
-50°…250°C
Ni 1 000/10 000 s 6 180 ppm/°C
-50°…250°C
Connection
2, 3 or 4-wire
Cu
Type
menu selectable
Cu 50/100 with 4 260 ppm/°C
-50°…200°C
Cu 50/100 with 4 280 ppm/°C
-200°…200°C
Connection
2-, 3-, or 4-wire
T/C
Type
selecmenu selectable
K (NiCr-Ni)
-200°…1 300°C
E (NiCr-CuNi)
-200°…690°C
B (PtRh30-PtRh6)
300°…1 820°C
S (PtRh10-Pt)
-50°…1 760°C
R (Pt13Rh-Pt)
-50°…1 740°C
N (Omegalloy)
-200°…1 300°C
DU
Range
Configuration is performed in two steps:
1.
|
setting numerical value for the beginning and end of the potentiometer range
|
2.
|
slider position calibration for the beginning and end of the potentiometer runway
|
Potentiometer power supply
2 VDC/6 mA, potentiometer resistance > 500 Ohm
External inputs
Number
3 inputs, on contact
Functions
OFF
|
No function
|
HOLD
|
Stopping measurement (option: Display - Analog output - Limits)
|
TARA
|
Tare activation
|
NUL. TA.
|
Resetting tare
|
NUL. M.M.
|
Resetting Min/Max. value
|
Projection
Display range
-999…9999, 3-coloured 7-segment LED
-99999…999999, 3-coloured 7-segment LED
-999…9999, single-coloured intensive LED (1200 mcd)
-99999…999999, single-coloured intensive LED (1200 mcd)
Number of characters
4 characters (only for height 100/125 mm
6 characters
Display height
57 mm, 100 mm, 125 mm
Display colour
red, green, red/green/orange
Signalizing LED
1, 2, 3, 4
|
Limits L1, L2, L3, L4 (red)
|
T
|
Tare (green)
|
M
|
Min./Max. value (green)
|
Description
the last two digits of the display can be used to display measured variables (only 6-digit display)
Decimal point
menu adjustable - fixed or floating
Display brightness
menu adjustable
Displayed values
Kan A
|
Value from channel A
|
Fil. A
|
Value from channel A after digital filter adjustment
|
Mat. Fn.
|
Value of mathematical function
|
Max.
|
Maximum value
|
Min.
|
Minimum value
|
Settings
Control
The instrument is set and controlled by an IR remote control
Selected functions can also be controlled via external inputs
Menu
LIGHT
|
contains only items necessary for basic device settings with password access
|
PROFI
|
Full device settings with password access
|
USER
|
User menu of any item, selected from LIGHT/PROFI. Access without password.
|
Optional functions
Remote control buttons "Left", "Up", "Down", "Enter" can be assigned additional optional features
ZAKAZ
|
button has no function
|
NUL. M.M.
|
Resetting Min/Max vaslues
|
NUL. TA.
|
Tare reset
|
MENU
|
Direct access to the selected menu item
|
DOC. H.
|
Temporary projection of selected values
|
OM Link
Company communication interface for operation, setting and update of device firmware
OM Link program is also designed for visualization and filing of meas. values from more instruments
Instrument accuracy and functionality
Temperature coefficient
50 ppm/°C
Accuracy
±0,1 % of range + 1 digit
±0,15 % of range + 1 digit (RTD, T/C)
(for projection 9999 and 5 measur./s)
Accuracy of cold junction measurement
±1,5°C
Overload capacity
2x; 10x (t < 30 ms)
Resolution (RTD, T/C)
1°C, 0,1°C, 0,01°C
Wire compensation (RTD, OHM)
< 30 Ohm
Cold junction compensation
adjustable -20°…99°C, automatic
Linearization
linear interpolation in 50 points (only via OM Link)
Digital filters
arithmetic mean
|
from 2…100 measurements
|
floating mean
|
from 2…30 measurements
|
exponential mean
|
from 2…100 measurements
|
rounding
|
setting projection step for display
|
Functions
Tare
|
designed to reset display on non-zero input signal
|
Min/max value
|
registration of min./max. value achieved during measurement
|
Peak value
|
display shows only max. or min. values
|
Math. functions
polynom, 1/x, logarithm, exponential, power, root
Math. operations between inputs
sum and quotient
Watch-dog
reset after 400 ms
Calibration
at 25°C and 40 % r.h.
Comparators
Type
Digital, menu adjustable, contact connect. < 30 ms
Mode
Hysteresis
|
switch. point, hysteresis band "Mez ± 1/2 Hys." and time (0 ... 99.9 s) determ. switching delay
|
From-To
|
output interval ON and OFF
|
Dose
|
period, its multiples and time (0 ... 99.9 s), during which the output is active
|
Output
1x relay, switch on contact (250 VAC/30 VDC, 3A)
2x relay, switch on contact (250 VAC/30 VDC, 3 A)
3x relay, switch on contact (250 VAC/30 VDC, 3A)
4x relay, switch on contact (250 VAC/30 VDC, 3A)
Data outputs
Protocol
ASCII, MESSBUS, MODBUS - RTU, PROFIBUS DP
Data format
8 bits + no parity + 1 stop bit (ASCII)
7 bits + even parity + 1 stop bit (Messbus)
Rate
600…230 400 Baud
0,0096…12 MBaud (PROFIBUS)
RS 485
isolated, addressing max. 31 instruments
Analogue outputs
Type
Isolated, programmable with a 16-bit D / A converter, type and range are menu selectable
Temperature coefficient
15 ppm/°C
Nonlinearity
±0,1 % of range
Rate
response to value change < 1 ms
Ranges
0...2 V, 0...5 V, 0...10 V, ±10 V, 0...5 mA, 0...20 mA, 4...20 mA
Compensation
600 Ohm/12 V
Error state
at range of 4...20mA
error message indication (<3.0 mA)
Detection of current loop interruption
Sensor excitation
Adjustable
5…24 VDC/max. 1,2 W
Power supply
Power supply
10…30 V AC/DC, ±10 %, PF > 0,4, I < 45 A/1 ms, isolated
80…250 V AC/DC, ±10 %, PF > 0,4, I < 45 A/2 ms, isolated
Current draw
< 22 W/22 VA
Power supply protection
Power supply is protected by a fuse inside the instrument
Operating conditions
Connection
cable bushings
connector terminal board, cable section < 1,5/2,5 mm2
Stabilization period
within 15 minutes after switch on
Operating temperature
-20°…60°C
Storage temperature
-20°…80°C
El. safety
EN 61010-1, A2
Dielectric strength
4 kVAC within 1 min. between pow. supply and input
4 kVAC within 1 min. between pow. supply and data/anal. output
4 kVAC within 1 min. between input and relay output
2,5 kVAC within 1 min.between input and data/anal. output
Insulation resistance
for polution degree II, measuring cat. III.
power supply
|
670 V (ZI), 300 V (Di)*
|
input, output, PN
|
300 V (ZI), 150 V (Di)*
|
Mechanical properties
Material
Anodized aluminum, black
Dimensions
57 mm/6 d.
|
375 x 119 x 88 mm
|
100 mm/4 d.
|
465 x 181 x 88 mm
|
100 mm/6 d.
|
651 x 181 x 88 mm
|
125 mm/4 d.
|
539 x 237 x 88 mm
|
125 mm/6 d.
|
754 x 237 x 88 mm
|
Panel cutout
57 mm/6 d.
|
367 x 111 mm
|
57 mm/6 d.
|
367 x 111 mm
|
100 mm/6 d.
|
643 x 173 mm
|
125 mm/4 d.
|
531 x 228 mm
|
125 mm/6 d.
|
746 x 228 mm
|
Installation
Into panel
On the wall/ceiling using bracket (included)
Images
Dimensions and installation
{"ds01-6949-2015.1-en":"11617","ds01-6949-2015.1-ru":"11618","ds01-6949-2015.1-cs":"11629","ds01-6949-2014.2-en":"11614","ds01-6949-2014.2-fr":"11615","ds01-6949-2014.2-ru":"11616","ds01-6949-2014.2-cs":"11628","ds01-6949-2014.1-en":"11611","ds01-6949-2014.1-fr":"11612","ds01-6949-2014.1-ru":"11613","ds01-6949-2014.1-cs":"11627","ds01-6949-2013.2-en":"11608","ds01-6949-2013.2-fr":"11609","ds01-6949-2013.2-ru":"11610","ds01-6949-2013.2-cs":"11626","ds01-6949-2013.1-ru":"11607","ds01-6949-2012.2-en":"11604","ds01-6949-2012.2-fr":"11605","ds01-6949-2012.2-ru":"11606","ds01-6949-2012.2-cs":"11625","ds01-6949-2010.1-en":"11601","ds01-6949-2010.1-fr":"11602","ds01-6949-2010.1-ru":"11603","ds01-6949-2010.1-cs":"11623","ds01-6949-2023.1-en":"10197","ds01-6949-2023.1-cs":"10198","ds01-6949-2020.1-en":"11597","ds01-6949-2020.1-fr":"11598","ds01-6949-2020.1-ru":"11599","ds01-6949-2020.1-cs":"11622","ds01-6949-2019.1-en":"11594","ds01-6949-2019.1-fr":"11595","ds01-6949-2019.1-ru":"11596","ds01-6949-2019.1-cs":"11621","ds01-6949-2017.1-en":"11592","ds01-6949-2017.1-ru":"11593","ds01-6949-2017.1-cs":"11620","ds01-6949-1-fr":"11600","ds01-6949-2012.1-cs":"11624","ds01-6949-2016.1-cs":"11619","man0100-6950-2020.3-en":"723","man0100-6950-2020.3-cs":"722","man01-6950-2021.4.2-en":"5410","man01-6950-2021.4.2-cs":"5405","man01-6950-2021.4.1-en":"5411","man01-6950-2021.4.1-cs":"5406","man0101-6950-2014.2.1-en":"742","man0101-6950-2014.2.1-cs":"737","man0101-6950-2014.2.0-cs":"736","man0102-6950-2010.1.2-en":"749","man0102-6950-2010.1.2-cs":"747","man0102-6950-2007.1.0-cs":"746","man0102-6950-2010.1.1-en":"748","man05-6950-2019.4.0-en":"5447","man05-6950-2019.4.0-cs":"5446","man06-6950-2018.4.0-en":"5455","man06-6950-2018.4.0-cs":"5449","man06-6950-2018.4.2-en":"5452","man06-6950-2018.4.2-cs":"5448","man06-6950-2018.4.1-en":"5454","man06-6950-2018.4.1-cs":"5450","man07-6950-2014.1.0-ru":"5457","man08-6950-2013.3.0-ru":"5458","man09-6950-2012.3.0-en":"5460","man09-6950-2012.3.0-cs":"5459","man010-6950-2011.1.1-en":"5461","an05-6951-2015-en":"5442","an05-6951-2015-cs":"5438","an05-6951-2012-en":"5439","an05-6951-2012-cs":"5435","an05-6951-2013-en":"5440","an05-6951-2013-cs":"5436","an05-6951-2014-en":"5441","an05-6951-2014-cs":"5437","an05-6951-2006.01-en":"1496","an05-6951-2021.01-cz":"1497","crt02-6952-2022-en":"13269","crt02-6952-2022-cs":"13270","crt03-6952-2022-en":"13073","crt03-6952-2022-cs":"13074","fw01-6953-2019-xx":"5409","sw03-6954-2020-xx":"752","sw04-6954-2020.2.33-xx":"755","sw05-6954-2024.2.34-xx":"13748"}
ISOLATED TRANSDUCER • USB < > OM INSTRUMENTS
Small compact transducer OM LINK USB with cable ensures safe and comfortable galvanically isolated connection between your computer and ORBIT MERRET instruments.